Human Membrane Protein, Palmitoylated 2 (MPP2) ELISA Kit
The Human Membrane Protein Palmitoylated 2 (MPP2) ELISA Kit is a cutting-edge tool designed for the precise detection of MPP2 levels in human samples, including serum, plasma, and cell culture supernatants. With its superior sensitivity and specificity, this kit provides accurate and reproducible results, making it ideal for a variety of research applications.MPP2 is a vital membrane protein that plays a crucial role in various cellular processes, including cell signaling and membrane trafficking.
Dysregulation of MPP2 has been linked to numerous diseases, making it a valuable biomarker for studying these conditions and developing potential therapeutic interventions.With the Human Membrane Protein Palmitoylated 2 (MPP2) ELISA Kit, researchers can delve deeper into the role of MPP2 in health and disease, paving the way for advancements in biomedical research and clinical diagnostics.
Product Name:
Human Membrane Protein, Palmitoylated 2 (MPP2) ELISA Kit
Product Code:
HUDL01898
Size:
96 Assays
Target:
MPP2
Synonyms:
DLG2, MAGUK P55 Subfamily Member 2, Discs Large,Homolog 2
Detection Method:
Sandwich
Range:
0.625-40ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.