Human Matrix Metalloproteinase 15 (MMP15) ELISA Kit
The Human Matrix Metalloproteinase 15 (MMP15) ELISA Kit is specifically designed for the accurate detection of MMP15 levels in human samples such as serum, plasma, and cell culture supernatants. This kit is known for its high sensitivity and specificity, ensuring precise and consistent results that are essential for various research applications.MMP15 is a key enzyme belonging to the matrix metalloproteinase family, playing a crucial role in extracellular matrix remodeling and tissue repair. Abnormal levels of MMP15 have been associated with various diseases such as cancer, inflammatory disorders, and cardiovascular conditions, highlighting its significance as a potential therapeutic target and biomarker for disease progression.
By using the Human MMP15 ELISA Kit, researchers can investigate the role of MMP15 in disease pathogenesis, monitor treatment response, and potentially discover new therapeutic strategies. With its reliable performance and easy-to-use protocol, this kit is a valuable tool for advancing research in the field of matrix metalloproteinases and their implications in human health.
Product Name:
Human Matrix Metalloproteinase 15 (MMP15) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-103
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.