The Human Macrophage Inflammatory Protein 3 Alpha (MIP-3a) ELISA Kit is a highly sensitive and specific assay designed for the accurate detection of MIP-3a levels in human serum, plasma, and cell culture supernatants. This kit is ideal for researchers studying inflammation, immune responses, and chemotaxis.MIP-3a is a chemokine that plays a key role in recruiting immune cells to sites of inflammation and infection. It is involved in regulating the migration and activation of various immune cells, making it an important biomarker for understanding the immune response and developing potential therapies for inflammatory diseases.
With its reliable and reproducible results, the Human MIP-3a ELISA Kit is a valuable tool for a wide range of research applications in immunology, inflammation, and infectious diseases. By accurately measuring MIP-3a levels, researchers can gain insights into the mechanisms underlying immune responses and inflammation, paving the way for new discoveries and treatments.
Product Name:
Human Macrophage Inflammatory Protein 3 Alpha (MIP3a) ELISA Kit
Product Code:
HUDL01858
Size:
96 Assays
Target:
MIP3a
Synonyms:
CCL20, CKb4, LARC, MIP3-A, SCYA20, ST38, Small Inducible Cytokine Subfamily A(Cys-Cys)Member 20, Beta-Chemokine Exodus-1, Liver And Activation-Regulated Chemokine
Detection Method:
Sandwich
Range:
78.125-5,000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
91-98
94
Heparin plasma (n=5)
90-99
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.