The Human Lymphocyte Antigen-9 (LY9) ELISA Kit is a highly sensitive and specific assay designed for the precise measurement of LY9 levels in human serum, plasma, and cell culture supernatants. This kit is perfect for researchers looking to study the role of LY9 in immune responses and autoimmune diseases.LY9 is a cell surface protein that plays a critical role in regulating lymphocyte activation and immune response. Dysregulation of LY9 has been linked to various autoimmune disorders and immunodeficiency diseases, highlighting its importance as a potential biomarker for disease diagnosis and therapeutic development.
With its accurate and reproducible results, the Human LY9 ELISA Kit is an essential tool for researchers studying lymphocyte biology and immune system function. By measuring LY9 levels, scientists can gain valuable insights into the mechanisms underlying immune-related disorders and explore new treatment strategies.
Product Name:
Human Lymphocyte Antigen 9 (LY9) ELISA Kit
Product Code:
HUDL01772
Size:
96 Assays
Target:
LY9
Synonyms:
RASI-1, MMP18, Occasionally Referred To As Stromelysin-4
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.