The Human Leukotriene A4 Hydrolase (LTA4H) ELISA Kit is specifically designed for the quantitative detection of LTA4H levels in human serum, plasma, and tissue lysates. This kit offers exceptional sensitivity and specificity, ensuring accurate and reliable results for a variety of research applications.LTA4H is an enzyme that plays a crucial role in the inflammatory response and has been implicated in various disease conditions, including asthma, arthritis, and inflammatory bowel disease.
By measuring LTA4H levels, researchers can gain valuable insights into the underlying mechanisms of these diseases and potentially identify new therapeutic targets.With its easy-to-use protocol and high-quality components, the Human LTA4H ELISA Kit is a valuable tool for researchers studying inflammation, immunology, and drug development. Trust in this kit to deliver precise and consistent results for your research needs.
Product Name:
Human Leukotriene A4 Hydrolase (LTA4H) ELISA Kit
Product Code:
HUDL01764
Size:
96 Assays
Target:
LTA4H
Synonyms:
LTA4 Hydrolase
Detection Method:
Sandwich
Range:
0.312-20ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
91-98
94
Heparin plasma (n=5)
90-99
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.