Human Growth Factor Receptor Bound Protein 10 (Grb10) ELISA Kit
The Human Growth Factor Receptor Bound Protein 10 (GRB10) ELISA Kit is specifically designed for the quantitative detection of GRB10 levels in human serum, plasma, and cell culture supernatants. With a high level of sensitivity and specificity, this kit provides accurate and reproducible results, making it an essential tool for various research applications.GRB10 is a key protein that plays a crucial role in cell growth and proliferation, as well as in insulin signaling pathways. Dysregulation of GRB10 has been linked to various diseases, including cancer and diabetes, making it a critical biomarker for studying these conditions and developing potential therapeutic interventions.
By using the Human GRB10 ELISA Kit, researchers can gain valuable insights into the role of GRB10 in disease pathogenesis and progression, ultimately leading to advancements in diagnosis and treatment options. Upgrade your research capabilities today with this cutting-edge ELISA kit from AssayGenie.
Product Name:
Human Growth Factor Receptor Bound Protein 10 (Grb10) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
90-103
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.