The Human Glucosidase Beta Acid (GBA) ELISA Kit is a highly sensitive and specific assay designed for the accurate detection of GBA levels in human biological samples such as serum, plasma, and cell culture supernatants. This kit provides reliable and reproducible results, making it an ideal tool for a wide range of research applications.GBA is an essential enzyme involved in the breakdown of glycolipids, and its dysfunction is linked to various lysosomal storage diseases such as Gaucher disease.
Monitoring GBA levels can provide valuable insights into lysosomal function and aid in the diagnosis and treatment of related disorders. With its high-performance features, the Human GBA ELISA Kit is a valuable resource for researchers studying lysosomal storage diseases, neurodegenerative disorders, and other conditions where GBA plays a critical role. It enables precise quantification of GBA levels, allowing for detailed analysis of disease mechanisms and potential therapeutic interventions.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-99
91
EDTA plasma (n=5)
80-98
89
Heparin plasma (n=5)
82-99
90
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.