Human Free Fatty Acid Receptor 1 (FFAR1) ELISA Kit
The Human Free Fatty Acid Receptor 1 (FFAR1) ELISA Kit is a powerful tool for the accurate measurement of FFAR1 levels in human serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring precise and reliable results for a variety of research applications.FFAR1, also known as GPR40, is a G protein-coupled receptor that binds to free fatty acids and plays a crucial role in regulating glucose and lipid metabolism.
Dysregulation of FFAR1 has been linked to metabolic disorders such as obesity, diabetes, and cardiovascular disease, making it a valuable biomarker for research and potential therapeutic interventions.With its high performance and user-friendly design, the Human FFAR1 ELISA Kit is an essential tool for scientists studying the role of FFAR1 in health and disease. Get accurate and reproducible results with this cutting-edge ELISA kit from Assay Genie.
Product Name:
Human Free Fatty Acid Receptor 1 (FFAR1) ELISA Kit
Product Code:
HUDL01047
Size:
96 Assays
Target:
FFAR1
Synonyms:
GPR40, FFA1R, G Protein-Coupled Receptor 40
Detection Method:
Sandwich
Range:
1.56-100ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.