Human FMS Like Tyrosine Kinase 3 Ligand (Flt3L) ELISA Kit
The Human FMS-like tyrosine kinase 3 ligand (FLT3L) ELISA Kit is a valuable tool for accurately measuring FLT3L levels in human samples such as serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring precise and consistent results for various research applications.FLT3L is a key cytokine that plays a crucial role in regulating the development and function of immune cells, particularly dendritic cells. Dysregulation of FLT3L levels has been implicated in various diseases, including cancer, autoimmune disorders, and infectious diseases.
Therefore, the Human FLT3L ELISA Kit is essential for studying the role of FLT3L in these conditions and developing potential therapeutic interventions.Whether you are investigating immune responses, cancer progression, or potential treatment options, the Human FLT3L ELISA Kit provides reliable and accurate data to advance your research efforts. Trust in this kit to help you uncover valuable insights into FLT3L biology and its impact on human health. Visit www.assaygenie.com/human-fms-like-tyrosine-kinase-3-ligand-flt3l-elisa-kit/ to learn more and order yours today.
Product Name:
Human FMS Like Tyrosine Kinase 3 Ligand (Flt3L) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
91-98
94
Heparin plasma (n=5)
90-99
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.