The Human Ferrochelatase (FECH) ELISA Kit from Assay Genie is specifically designed for the sensitive and accurate detection of FECH levels in human samples including serum, plasma, and cell culture supernatants. This kit is highly reliable and reproducible, providing researchers with consistent and precise results for a variety of research applications.Ferrochelatase is a key enzyme involved in heme biosynthesis, playing a crucial role in the binding of iron to protoporphyrin IX to form heme. Dysregulation of FECH has been linked to various disorders such as porphyria and iron overload syndromes, making it an important biomarker for studying these conditions and developing potential therapeutic interventions.
By utilizing the Human Ferrochelatase ELISA Kit, researchers can accurately quantify FECH levels in human samples, allowing for a better understanding of the role of this enzyme in various physiological and pathological processes. With its high sensitivity and specificity, this kit is an invaluable tool for advancing research in the fields of hematology, biochemistry, and molecular biology.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.