Human Endothelin Converting Enzyme 1 (ECE1) ELISA Kit
The Human Endothelin-Converting Enzyme 1 (ECE1) ELISA Kit is specifically designed for the precise measurement of ECE1 levels in human serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring dependable and consistent results for various research applications.Endothelin-Converting Enzyme 1 is a crucial enzyme involved in the regulation of endothelin, a potent vasoconstrictor that plays a key role in the cardiovascular system. Dysregulation of ECE1 has been implicated in various cardiovascular diseases, including hypertension and atherosclerosis.
Therefore, measuring ECE1 levels can provide valuable insights into the pathophysiology of these conditions and aid in the development of potential therapeutic interventions.Overall, the Human ECE1 ELISA Kit is an indispensable tool for researchers studying cardiovascular diseases and related disorders, offering a reliable and efficient method for quantifying ECE1 levels in biological samples.
Product Name:
Human Endothelin Converting Enzyme 1 (ECE1) ELISA Kit
Product Code:
HUDL00905
Size:
96 Assays
Target:
ECE1
Detection Method:
Sandwich
Range:
31.25-2000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.