The Human Endocrine Gland-Derived Vascular Endothelial Growth Factor (EG-VEGF) ELISA Kit is specially designed for the accurate quantification of EG-VEGF levels in human serum, plasma, and cell culture supernatants. This ELISA kit offers high sensitivity and specificity, ensuring precise and reproducible results for various research purposes.EG-VEGF is a vital protein involved in vascular development, particularly in endocrine glands, and plays a crucial role in regulating angiogenesis and vascular permeability. Dysregulation of EG-VEGF has been implicated in various diseases, including cancer, endocrine disorders, and reproductive disorders, making it a valuable biomarker for studying these conditions and exploring potential therapeutic interventions.
Overall, the Human Endocrine Gland-Derived Vascular Endothelial Growth Factor (EG-VEGF) ELISA Kit provides researchers with a reliable tool for studying the role of EG-VEGF in health and disease, paving the way for advancements in the field of vascular biology and therapeutic development.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-101
91
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.