The Human E1A-binding protein p300 (EP300) ELISA Kit is specifically designed for the accurate detection of E1A-binding protein p300 levels in human serum, plasma, and cell culture supernatants. This kit is highly sensitive and specific, providing reliable and reproducible results for a wide range of research applications.E1A-binding protein p300, also known as EP300, plays a crucial role in regulating gene expression, cell growth, and differentiation. It is involved in various cellular processes, including DNA repair, apoptosis, and cell cycle control.
Dysregulation of EP300 has been linked to cancer, developmental disorders, and cardiovascular diseases, making it a valuable biomarker for studying these conditions and developing potential treatments.The Human E1A-binding protein p300 (EP300) ELISA Kit is a valuable tool for researchers studying the role of EP300 in disease development and progression. Its high sensitivity and specificity make it a reliable option for accurately measuring E1A-binding protein p300 levels in biological samples.
Product Name:
Human E1A Binding Protein P300 (EP300) ELISA Kit
Product Code:
HUDL00953
Size:
96 Assays
Target:
EP300
Synonyms:
KAT3B, p300, Histone Acetyltransferase p300
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
83-100
91
EDTA plasma (n=5)
91-102
96
Heparin plasma (n=5)
89-102
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.