The Human Cytochrome P450 27B1 (CYP27B1) ELISA Kit is specifically designed for the accurate quantification of CYP27B1 levels in human serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring precise and reliable results for a variety of research applications.Cytochrome P450 27B1 is a key enzyme involved in the activation of vitamin D, playing a crucial role in calcium metabolism and bone health. Dysregulation of CYP27B1 has been associated with various diseases, including osteoporosis, autoimmune disorders, and certain cancers, making it a valuable biomarker for understanding these conditions and developing potential therapeutic interventions.
With its superior performance and ease of use, the Human Cytochrome P450 27B1 (CYP27B1) ELISA Kit is an indispensable tool for researchers studying vitamin D metabolism and related diseases. Trust in this kit to deliver accurate and reproducible data for advancing your research efforts.
Product Name:
Human Cytochrome P450 27B1 (CYP27B1) ELISA Kit
Product Code:
HUDL00766
Size:
96 Assays
Target:
CYP27B1
Synonyms:
VDD1, PDDR, CYP1, P450c1, 25-Hydroxyvitamin D3 1-Alpha-Hydroxylase, Cytochrome P450 Family 27 Subfamily B Polypeptide 1, Calcidiol 1-monooxygenase, VD3 1A hydroxylase
Detection Method:
Sandwich
Range:
0.78-50ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-101
91
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.