Human Cholecystokinin 8, Octapeptide (CCK8) ELISA Kit
The Human Cholecystokinin-8 (CCK8) Octapeptide ELISA Kit is a powerful tool for the precise measurement of CCK8 levels in human samples such as serum, plasma, and cell culture supernatants. This kit offers exceptional sensitivity and specificity, ensuring accurate and reproducible results for a variety of research purposes.Cholecystokinin-8 (CCK8) is a key peptide involved in regulating appetite, digestion, and the release of digestive enzymes. It plays a crucial role in the gastrointestinal system and has been implicated in disorders such as obesity, eating disorders, and gastrointestinal diseases.
As such, CCK8 serves as a valuable biomarker for investigating these conditions and developing potential treatments.With its user-friendly design and reliable performance, the Human Cholecystokinin-8 (CCK8) Octapeptide ELISA Kit is an essential tool for researchers studying the role of CCK8 in physiology and disease. Trust in this kit to deliver accurate and insightful results that can advance your research efforts.
Product Name:
Human Cholecystokinin 8, Octapeptide (CCK8) ELISA Kit
Product Code:
HUDL00455
Size:
96 Assays
Target:
CCK8
Synonyms:
Sincalide
Detection Method:
Competition
Range:
15.625-1000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.