Human ATP Binding Cassette Transporter C6 (ABCC6) ELISA Kit
The Human ATP Binding Cassette Transporter C6 (ABCC6) ELISA Kit is a powerful tool for the precise quantification of ABCC6 levels in human biological samples, including serum, plasma, and cell culture supernatants. With its exceptional sensitivity and specificity, this kit ensures accurate and consistent results, making it suitable for a wide range of research purposes.ABCC6 is a vital transporter protein involved in the ATP-binding cassette (ABC) transporter superfamily, playing a crucial role in transporting various substances across cell membranes.
Dysregulation of ABCC6 has been linked to genetic disorders such as pseudoxanthoma elasticum (PXE), highlighting its importance in normal physiological functions.By accurately measuring ABCC6 levels, researchers can gain valuable insights into its role in different cellular processes and diseases, paving the way for potential therapeutic strategies targeting ABCC6-related conditions. Explore the potential of the Human ABCC6 ELISA Kit for your research needs.
Product Name:
Human ATP Binding Cassette Transporter C6 (ABCC6) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.