Human Amine Oxidase, Copper Containing 3 (AOC3) ELISA Kit
The Human Amine Oxidase Copper-Containing 3 (AOC3) ELISA Kit is a highly sensitive and specific assay designed for the accurate measurement of AOC3 levels in human serum, plasma, and cell culture supernatants. AOC3 is an important enzyme involved in the regulation of amine metabolism, impacting various physiological processes.This kit offers reliable and reproducible results, making it perfect for a wide range of research applications related to metabolism, inflammation, and neurological disorders.
AOC3 levels have been linked to conditions such as diabetes, Alzheimer's disease, and cardiovascular disease, showcasing its potential as a valuable biomarker for disease research and therapeutic development.Order the Human Amine Oxidase Copper-Containing 3 (AOC3) ELISA Kit today to advance your research and gain valuable insights into the role of this enzyme in human health and disease.
Product Name:
Human Amine Oxidase, Copper Containing 3 (AOC3) ELISA Kit
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.