Guinea pig Regulated On Activation In Normal T-Cell Expressed And Secreted (RANTES) ELISA Kit
The Guinea Pig Regulated On Activation, Normal T Cell Expressed, and Secreted (RANTES) ELISA Kit is a powerful tool for detecting RANTES levels in guinea pig serum, plasma, and cell culture supernatants. This kit offers superior sensitivity and specificity, ensuring accurate and consistent results for a variety of research applications.RANTES is a key chemokine involved in the immune response, playing a role in inflammation and recruitment of immune cells to sites of infection or injury.
Dysregulation of RANTES has been implicated in various inflammatory conditions, autoimmune diseases, and viral infections, making it a valuable biomarker for studying these diseases and potential therapeutic interventions.Overall, the Guinea Pig RANTES ELISA Kit provides researchers with a reliable and efficient method for quantifying RANTES levels in guinea pig samples, offering insights into the role of this chemokine in immune responses and disease pathology.
Product Name:
Guinea pig Regulated On Activation In Normal T-Cell Expressed And Secreted (RANTES) ELISA Kit
Product Code:
GPDL00023
Size:
96 Assays
Target:
RANTES
Synonyms:
CCL5, SCYA5, SISd, EoCP, Chemokine C-C-Motif Ligand 5, Eosinophil Chemotactic Cytokine, T-cell Specific Protein p288, Small Inducible Cytokine Subfamily A(Cys-Cys)Member 5
Detection Method:
Sandwich
Range:
0.156-10ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.