The Goat Interleukin-10 (IL-10) ELISA Kit is a reliable and sensitive assay designed for the accurate measurement of IL-10 levels in goat serum, plasma, and cell culture supernatants. This kit offers high specificity and sensitivity, providing researchers with precise and reproducible results for a variety of applications.IL-10 is a key regulatory cytokine that plays a critical role in modulating immune responses and inflammation. Dysregulation of IL-10 has been associated with various diseases, including autoimmune disorders, infectious diseases, and cancer.
By accurately quantifying IL-10 levels, researchers can gain valuable insights into the immune response and potential therapeutic targets for these conditions.The Goat IL-10 ELISA Kit is an essential tool for studying the role of IL-10 in goat physiology and pathology, offering researchers a reliable and efficient method for quantifying IL-10 levels in biological samples. With its high sensitivity and specificity, this kit provides accurate and reproducible results, making it an ideal choice for a wide range of research applications.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.