The General Histamine (HIS) ELISA Kit is specifically designed for the precise quantification of histamine levels in various biological samples including serum, plasma, and tissue culture supernatants. This kit boasts exceptional sensitivity and specificity, ensuring accurate and consistent results for a diverse array of research purposes.Histamine is a key biogenic amine that serves important functions in the body, such as regulating immune responses, controlling gastric acid secretion, and influencing neurological processes.
Abnormal histamine levels have been implicated in a range of conditions including allergies, asthma, and gastrointestinal disorders, making it a valuable biomarker for investigating these diseases and exploring potential therapeutic interventions.By utilizing the General Histamine (HIS) ELISA Kit, researchers can gain valuable insights into histamine levels and their role in various biological processes, advancing our understanding of histamine-related diseases and paving the way for novel treatment strategies.
Product Name:
General Histamine (HIS) ELISA Kit
Product Code:
UNDL00045
Size:
96 Assays
Target:
HIS
Synonyms:
Histamin
Detection Method:
Competition
Range:
3.12-200ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.