The General Epinephrine (Epi) ELISA Kit is specifically designed for the precise measurement of epinephrine levels in various biological samples including serum, plasma, and urine. This kit offers exceptional sensitivity and specificity, ensuring accurate and consistent results for a diverse range of research purposes.Epinephrine, also known as adrenaline, is a crucial hormone and neurotransmitter responsible for the body's fight or flight response.
It plays a vital role in regulating heart rate, blood pressure, and metabolism, making it a key biomarker in studying conditions such as stress, anxiety, and cardiovascular diseases.With its high-quality components and user-friendly protocol, the General Epinephrine (Epi) ELISA Kit is an invaluable tool for researchers looking to delve into the intricate mechanisms of epinephrine regulation and its impact on human health.
Product Name:
General Epinephrine (EPI) ELISA Kit
Product Code:
UNDL00035
Size:
96 Assays
Target:
EPI
Synonyms:
Adrenaline
Detection Method:
Competition
Range:
31.25-2000pg/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.