The Chicken Netrin-1 (NTN1) ELISA Kit is specifically designed for the accurate and reliable detection of netrin-1 levels in chicken serum, plasma, and cell culture supernatants. This kit offers high sensitivity and specificity, providing researchers with precise and reproducible results for a variety of research applications.Netrin-1 is a key protein involved in axon guidance and neuronal development, playing a crucial role in the formation of neural circuits. Dysregulation of netrin-1 has been implicated in neurological disorders such as Alzheimer's disease and Parkinson's disease, making it a valuable biomarker for studying these conditions and potential therapeutic interventions.
The Chicken Netrin-1 (NTN1) ELISA Kit is a valuable tool for researchers looking to investigate the role of netrin-1 in chicken development and disease, offering a reliable and efficient method for measuring netrin-1 levels in biological samples.
Product Name:
Chicken Netrin 1 (Ntn1) ELISA Kit
Product Code:
CHDL00047
Size:
96 Assays
Target:
Ntn1
Synonyms:
NTN1L, Epididymis tissue protein Li 131P
Detection Method:
Competition
Range:
15.625-1,000pg/m
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
88-101
94
EDTA plasma (n=5)
91-103
97
Heparin plasma (n=5)
92-98
95
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.