The Bovine Vasoactive Intestinal Peptide (VIP) ELISA Kit is a cutting-edge tool for the precise measurement of VIP levels in bovine samples such as serum, plasma, and tissue culture supernatants. This kit offers unparalleled sensitivity and specificity, ensuring accurate and consistent results for a variety of research applications.VIP is a key neuropeptide with potent vasodilatory and anti-inflammatory properties, playing a vital role in the regulation of various physiological processes including immune response, smooth muscle relaxation, and neurotransmitter release.
Dysregulation of VIP has been linked to numerous diseases such as asthma, inflammatory bowel disease, and autoimmune disorders, underscoring its significance as a valuable biomarker for investigating these conditions and potential therapeutic interventions.With its advanced technology and superior performance, the Bovine VIP ELISA Kit is an indispensable tool for researchers seeking to unravel the complex role of VIP in health and disease, paving the way for innovative discoveries and treatment strategies.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. Note:To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity, and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 50µL standard or sample to each well.    And then add 50µL prepared Detection Reagent A immediately.    Shake and mix. Incubate 1 hour at 37°C
3.
Aspirate and wash 3 times
4.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
5.
Aspirate and wash 5 times
6.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
7.
Add 50µL Stop Solution. Read at 450 nm immediately.