The Bovine Myeloperoxidase (MPO) ELISA Kit from Assay Genie is a versatile and reliable tool for the accurate detection of MPO levels in bovine serum, plasma, and cell culture supernatants. With high sensitivity and specificity, this kit delivers precise and reproducible results, making it perfect for a wide range of research applications.MPO is a key enzyme involved in the innate immune response, playing a crucial role in the oxidative burst process. It is implicated in various inflammatory conditions and is considered a biomarker for a range of diseases, including cardiovascular disorders and autoimmune diseases.
By measuring MPO levels, researchers can gain valuable insights into the pathogenesis of these conditions and develop potential therapeutic interventions.Trust the Bovine Myeloperoxidase (MPO) ELISA Kit from Assay Genie for accurate and reliable detection of MPO levels in bovine samples, advancing your research in immunology, inflammation, and disease pathology.
Product Name:
Bovine Myeloperoxidase (MPO) ELISA Kit
Product Code:
BODL00125
Size:
96 Assays
Target:
MPO
Detection Method:
Sandwich
Range:
1.56-100ng/mL
Shelf Life:
12 months
Note:
For research use only
Recovery:
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.