The Bovine Matrix Metalloproteinase-3 (MMP3) ELISA Kit is a reliable and accurate tool for the quantitative detection of MMP3 levels in bovine samples, including serum, plasma, and tissue homogenates. With its high sensitivity and specificity, this kit provides precise and reproducible results, making it ideal for a variety of research applications in the field of matrix metalloproteinase biology.MMP3, also known as stromelysin-1, is a key enzyme involved in the breakdown of extracellular matrix components, playing a crucial role in tissue remodeling and inflammation.
Dysregulation of MMP3 activity has been associated with various diseases, including arthritis, cancer, and cardiovascular disorders, making it a valuable biomarker for studying these conditions and developing potential therapeutic interventions.Overall, the Bovine MMP3 ELISA Kit offers researchers a powerful tool for investigating the role of MMP3 in physiological and pathological processes, ultimately leading to advancements in the understanding and treatment of related diseases.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.