The Bovine Growth Hormone (GH) ELISA Kit is specifically designed for the precise measurement of bovine growth hormone levels in various sample types, including serum, plasma, and tissue culture supernatants. With its superior sensitivity and specificity, this kit delivers accurate and reproducible results, making it an essential tool for researchers in the field.Bovine growth hormone is a key protein that plays a critical role in regulating growth, metabolism, and milk production in cattle. Monitoring its levels can provide valuable insights into the health and performance of livestock, as well as aid in the development of strategies to optimize growth and productivity.
Whether studying animal nutrition, reproductive physiology, or exploring the effects of growth hormone on muscle development, the Bovine Growth Hormone ELISA Kit offers a reliable and efficient solution for measuring bovine growth hormone levels in biological samples. Get your research on the fast track with this cutting-edge ELISA kit from Assay Genie.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
82-102
92
EDTA plasma (n=5)
90-100
95
Heparin plasma (n=5)
88-100
94
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.