The Bovine Fibroblast Growth Factor 1 (Acidic FGF1) ELISA Kit is a powerful tool for the accurate quantification of acidic FGF1 levels in bovine serum, plasma, and cell culture supernatants. With high sensitivity and specificity, this kit provides reliable and reproducible results, making it perfect for various research applications.Acidic FGF1, also known as fibroblast growth factor 1, is a key protein involved in cell growth and proliferation. It plays a critical role in various physiological processes, including wound healing, tissue regeneration, and embryo development.
Dysregulation of acidic FGF1 is associated with diseases such as cancer and metabolic disorders, highlighting its importance as a potential therapeutic target.With the Bovine Fibroblast Growth Factor 1 (Acidic FGF1) ELISA Kit, researchers can accurately measure acidic FGF1 levels, furthering our understanding of its biological functions and potential clinical implications. Get reliable results and unlock new insights with this advanced ELISA kit.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-99
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.