The Bovine Defensin Beta 2 (DEFB2) ELISA Kit is a high-quality assay designed for the accurate quantification of bovine defensin beta 2 levels in biological samples such as serum, plasma, and tissue homogenates. This kit offers exceptional sensitivity and specificity, ensuring precise and reproducible results for your research needs.Defensin beta 2 is an important antimicrobial peptide that plays a crucial role in the innate immune response of cattle. It helps protect against bacterial and viral infections by disrupting microbial cell membranes, making it a key molecule in host defense mechanisms.
The measurement of defensin beta 2 levels can provide valuable insights into the immune system function and disease pathology in bovine species.Whether you are studying infectious diseases, immune responses, or exploring new therapeutic strategies, the Bovine Defensin Beta 2 (DEFB2) ELISA Kit is an essential tool for your research. Trust in its accuracy and reliability to advance your understanding of bovine immunity and health.
Matrices listed below were spiked with certain level of recombinant the index and the recovery rates were calculated by comparing the measured value to the expected amount of the index in samples.
Matrix
Recovery range (%)
Average(%)
Serum (n=5)
80-102
91
EDTA plasma (n=5)
81-100
90
Heparin plasma (n=5)
80-89
84
Linearity:
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of the index and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
1:16
Serum (n=5)
82-96%
83-98%
81-99%
93-101%
EDTA plasma (n=5)
88-101%
86-95%
90-102%
80-93%
Heparin plasma (n=5)
80-91%
82-90%
95-104%
79-95%
Intra-assay Precision:
Intra-Assay: CV <10%. 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
Inter-assay Precision:
Inter-Assay: CV <12%. 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
Stability:
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
Step
Protocol
1.
Prepare all reagents, samples and standards
2.
Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3.
Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4.
Aspirate and wash 3 times
5.
Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6.
Aspirate and wash 5 times
7.
Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8.
Add 50µL Stop Solution. Read at 450nm immediately.