Bovine/Caprine/Sheep IL-5 DIY ELISA Kit
- SKU:
- KES0223
- Product Type:
- ELISA Kit
- Reactivity:
- Bovine
- Goat
- Sheep
- Applications:
- ELISA
- ELISA Type:
- DIY ELISA
- Size:
- 10 Plates
- Synonyms:
- IL5, EDF, IL-5, TRF
Description
Bovine/Caprine/Sheep IL-5 DIY ELISA Kit
The Bovine/Caprine/Ovine IL-5 ELISA Kit is specifically designed for the accurate and reliable detection of IL-5 levels in bovine, caprine, and ovine serum, plasma, and cell culture supernatants. This kit offers high sensitivity and specificity, ensuring trustworthy and consistent results for a variety of research applications.IL-5, also known as Interleukin-5, is a key cytokine involved in the regulation of the immune response, specifically in the development and differentiation of eosinophils. Dysregulation of IL-5 has been linked to inflammatory disorders, allergic diseases, and autoimmune conditions, making it an important target for research and potential therapeutic interventions.
With its precise and efficient detection capabilities, the Bovine/Caprine/Ovine IL-5 ELISA Kit is an essential tool for scientists and researchers studying immune responses and inflammatory diseases in bovine, caprine, and ovine models. Get reliable and reproducible results with this advanced ELISA kit from Assay Genie.
Product Name: | Bovine/Caprine/Sheep IL-5 DIY ELISA |
Product Code: | KES0223 |
Species: | Multi |
Target: | IL-5 |
Synonyms: | IL5, EDF, IL-5, TRF |
Application: | ELISA |
ELISA Type: | DIY ELISA Kit |
Size: | 10 Plates |
Shipping: | Room temperature |
Storage: | Stable for up to twelve months from date of receipt at 2-8°C. |
Description | Usage | Quantity |
Anti-Bovine/Caprine/Sheep IL-5 Polyclonal Antibody | Capture Antibody | 100 µg |
Biotinylated Anti-Bovine/Caprine/Sheep IL-5 Polyclonal Antibody | Detection Antibody | 50 µg |
Bovine/Caprine/Sheep IL-5 Recombinant Protein | Standard | 5 µg |
Reagent | Suggested Formulation |
DPBS: | 0.008M sodium phosphate, 0.002M potassium phosphate, 0.14M sodium chloride, 0.01M potassium chloride, pH 7.4 |
96-well ELISA Plate: | Clear, flat-bottom, high-binding 96-well plate, 8-wells per strip, 350 µL per well (ELISA Plates: KESAP003) |
Standard and Sample Diluent: | The optimal Standard and Sample Diluent will need to be determined for each sample type to obtain optimal recovery and linearity. The appropriate Standard and Sample Diluent will mimic the sample's response to a known quantity of protein standard and will provide linear results when diluted. Often a 1:4 dilution of the sample in Reagent Diluent will provide acceptable recovery and linearity. |
Reagent Diluent and Blocking Buffer: | 4% BSA in DPBS, 0.2 µm filtered |
Wash Buffer: | 0.05% Tween®-20 in DPBS |
Streptavidin-HRP: | Enzymatic reagent to react with biotinylated detection antibody (Streptavidin-HRP: KESAP002) |
Substrate: | 3,3',5,5'-tetramethylbenzidine (TMB) Substrate (ELISA Accessory Pack: KESAP001) |
Stop Solution: | 0.18 M Sulfuric Acid (ELISA Accessory Pack: KESAP001) |
Plate Sealer: | Adhesive film to prevent evaporation |
*Note: Protocols are specific to each batch/lot. For the correct instructions please follow the protocol included in your kit.
Step | Procedure |
1. | Prepare Capture Antibody in DPBS at desired working concentration. |
2. | Add 100 µL of Capture Antibody Working Solution to appropriate wells. |
3. | Cover plate with Plate Sealer and incubate at room temperature (20-25°C) for 12-24 hours. |
4. | Empty Capture Antibody Working Solution from plate. Blot plate onto paper towels or other absorbent material. |
5. | Add 100 µL of Blocking Buffer to appropriate wells. |
6. | Cover plate with Plate Sealer and incubate at room temperature for 1-3 hours. |
7. | Empty Blocking Buffer from plate. Blot plate onto paper towels or other absorbent material. |
8. | Prepare Standard and sample as desired with Standard and Sample Diluent. |
9. | Add 100 µL of Standard or sample to appropriate wells. Note: Run each Standard or sample in duplicate. |
10. | Cover plate with Plate Sealer and incubate at room temperature for 1 hour. |
11. | Wash plate FOUR times with Wash Buffer. Note: Gently squeeze the long sides of plate frame before washing to ensure all strips remain securely in the frame. Empty plate contents. Use a squirt wash bottle to vigorously fill each well completely with 1X Wash Buffer, then empty plate contents. Repeat procedure three additional times for a total of FOUR washes. Blot plate onto paper towels or other absorbent material. |
12. | Prepare Detection Antibody in Reagent Diluent at desired working concentration. |
13. | Add 100 µL of Detection Antibody Working Solution to each well. |
14. | Cover plate with Plate Sealer and incubate at room temperature for 1 hour. |
15. | Wash plate FOUR times with Wash Buffer as described in step 11. |
16. | Prepare Streptavidin-HRP in Reagent Diluent at desired working concentration. |
17. | Add 100 µL of Streptavidin-HRP Working Solution to each well. |
18. | Cover plate with Plate Sealer and incubate at room temperature for 30 minutes. |
19. | Wash plate FOUR times with Wash Buffer as described in step 11. |
20. | Add 100 µL of TMB Substrate Solution to each well. |
21. | Develop the plate in the dark at room temperature for 30 minutes or as desired. Note: Do NOT cover plate with Plate Sealer. |
22. | Stop reaction by adding 100 µL of Stop Solution to each well. |
23. | Measure absorbance on a plate reader at 450 nm. |
The Bovine/Caprine/Sheep IL-5 DIY ELISA kit from Assay Genie can assay for IL-5 in the following samples: serum, blood, plasma, cell culture supernatant and other related supernatants and tissues. Bovine/Caprine/Sheep IL-5 DIY ELISA Kit from Assay Genie allows researchers to develop their own ELISA plates for IL-5 using our unique combination of capture and detection antibodies.
Each Bovine/Caprine/Sheep IL-5 DIY ELISA Kit contains capture antibody, standard, and detection antibody for development of an IL-5 ELISA. IL-5 antibodies in this kit have been determined to function in an ELISA with the IL-5 standard provided. Optimal buffers, concentrations, incubation times, incubation temperatures, and methods for the ELISA have not been determined and require optimisation for the development of this kit. A working knowledge of ELISA is strongly recommended.